10 Allow more than 30 minutes for the unit to warm up so that it is stabilized and ready for use. 3. OPERATION 3-
11 the knob out and rotating it varies the level from negative to positive around the mid point. INPUT C, BNC : Input for frequency
12 e. TOTAL A. When this mode is selected, the unit counts cycles of the input A signal and continuously displays that count. f. T.l (AB). When
13 FIG (3) REAR PANEL AC INLET: AC power input connector VOLTAGE SELECTOR: Selects the AC Power (115 V or 230 V)
14 RS-232C CONNECTOR : Connector for serial interfacing with a computer. 3-2. Operating Instruction Below is the basic operatin
15 DC coupling position. 3-3-2 INPUT C (80 MHz to 2.4 GHz) CAUTION THE MAXIMUM INPUT LIMIT TO THIS INPUT IS 3 Vrms MAXIMUM OVER THE I
16 3-6. Time Interval Measurements(AB) In time interval mode, unit measures the elapsed time from a selected edge of the input A w
17 3-7. Frequency Ratio Measurements (A/B) In this mode of operation, the counter displays the ratio of the frequency applied to
18 3-8. Use of RS-232C Serial Interfacing 1) Hardware/Software Requirements . IBM PC/XT/AT or compatible computer . Microsoft Windo
19 4. MAINTENANCE CAUTION IT IS ESSENTIAL FOR SAFETY TO PROPERLY MAINTAIN AND SERVICE THIS INSTRUMENT WARNING VOLTAGES WITHIN THIS INSTRUME
2 Introduction Thank you for purchasing our product. Electronic measuring instruments produced by us are high technology products made un
20 5. OTHERS 5-1. BNC Cable Considerations Accuracy of radio frequency measurements can be affected by connections between signal sou
21 5-3. Line Frequency Measurements Use of the attenuator, low pass filter, and/or x10 probe is advisable when measuring line frequenc
22 Limited One-Year Warranty B&K Precision Corp. warrants to the original purchaser that its product and the component parts thereof, wil
23 Model Number: ______________ Date Purchased:__________ Service Information Warranty Service: Please re
24 PN: 481-542-9-001 Printed in Korea 2004 B&K Precision Corp. V4.8.2014 22820 Savi Ranch Parkway Yorba Lind
3 CONTENTS 1. PRODUCT DESCRIPTION 1-1. Introduction ------------------------------------------------------ ( 4 ) 1-2. Technical
4 1. PRODUCT DESCRIPTION 1-1. Introduction This reciprocal UNIVERSAL COUNTER series are microprocessor controlled instruments
5 1-2. Technical Specifications INPUT A CHARACTERISTICS FREQUENCY RANGE : 0.1 Hz to 100 MHz (DC Coupled) 30 Hz to 100 MHz (AC Coupled) S
6 MAX. INPUT VOLTAGE LEVEL FIG 1. MAX. Input Level. (Input A,B) INPUT B. CHARACTERISTICS = Same as Input A TI
7 INPUT C. CHARACTERISTICS FREQUENCY RANGE: 80 MHz to 2.4 GHz SENSITIVITY: 25 mV from 80 MHz to 150 MHz 20 mV from 150 MHz to 2.0 G
8 displayed. When Hold is released, a new measurement begins. GATE TIME : Depending on input frequency < 10 mS-----
9 Power cord-----------------------------------------------------------------1 Spare Fuse---------------------------------------------------