Bk-precision 2552-2559 - Programming Manual Manuel d'utilisateur Page 30

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 140
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 29
30
DESCRIPTION The CURSOR_VALUE? Query returns the
values measured by the specified cursors for a
given trace. (The PARAMETER_VALUE?
query is used to obtain measured waveform
parameter values.)
Notation
HREL
the cursor value under track cursor mode
VREL
the dalta volt-value under manual cursor mode
QUERY SYNTAX <trace>: CuRsor_Value? [<mode>,…<mode>]
<trace> : = { C1, C2, C3, C4}
<mode> : = { HREL, VREL }
RESPONSE FORMAT <trace> : CuRsor_Value HREL,
<B->T - A->T>,<B->V - A->V>,<A->T>,
<B->T>,
<(B->V - A->V)/(B->T - A->T)>
<trace> : CuRsor_Value VREL,<delta_vert>
EXAMPLE The following query reads the dalta volt value
under manual cursor mode (VREL) on
Channel 2:
Command message:
C2:CRVA? VREL
Response message:
C2:CuRsor_Value VREL 1.00V
RELATED COMMANDS CURSOR_SET, PARAMETER_VALUE
CURSOR
CURSOR_VALUE?, CRVA?
Query
Vue de la page 29