Bk-precision 2552-2559 - Programming Manual Manuel d'utilisateur Page 35

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 140
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 34
35
DESCRIPTION The DATE command changes the date/time of the
oscilloscope’s internal real-time clock.
The command is only used in the CFL series
instrument.
COMMAND SYNTAX DATE <day>, <month>, <year>, <hour>,
<minute>, <second>
<day> : = 1 to 31
<month> : = {JAN, FEB, MAR, APR, MAY,
JUN, JUL, AUG, SEP,OCT, NOV, DEC}
<year> : = 1990 to 2089
<hour> : = 0 to 23
<minute> : = 0 to 59
<second> : = 0 to 59
QUERY SYNTAX DATE?
RESPONSE FORMAT DATE <day>, <month>, <year>, <hour>,
<minute>, <second>
EXAMPLE This instruction will change the date to
NOV. 1, 2009 and the time to 14:38:16:
Command message:
DATE 1, NOV, 2009,14,38,16
MISCELLANEOUS
DATE
Command /Query
Vue de la page 34